Disable Preloader

Lorem ipsum is simply free text dolor sit am adipi we help you ensure everyone is in the right jobs sicing elit, sed do consulting firms Et leggings across the nation tempor.

Latest News

Contact Info
images

Dust Monitor [Nephelometer]

The Sensidyne Nephelometer is an advanced real-time dust monitor (aerosol monitor) accurately measuring dust concentrations using proven light scatter technology. This portable instrument accurately measures and records dust from 1-10,000 μg/m3 with resolution to 1 μg/m3. Sample modes are selectable between 60 second sample, 15 minute STEL, or continuous sampling.

The Nephelometer has an internal pump drawing samples into the iso-kinetic sampling inlet where they meet sheath air that guides samples past the particle sensor. The sensor is a photo detector that measures laser light scattered by particulates in the sample stream. The instrument multiplies each measurement by a K-factor and displays the real-time reading on the display of the instrument. After each sample, the K-factor and Environmental Factor Name and STEL, maximum, minimum, and average readings record to the internal data log. When connected to a computer the instrument uploads up to 4,000 data-log records in spreadsheet format.

In addition to high sensitivity and ease-of-use the Sensidyne Nephelometer offers a low cost of ownership, automatic power save, long-life, data-logging, facility monitoring features, and user-replaceable filters. The Sensidyne Nephelometer is a competitive, well priced alternative to high priced aerosol monitors.

 

Accurate Dust Concentration

Proprietary algorithm provides accurate dust concentration. The sheath air feature prevents internal contamination and improves accuracy.

 

Reliable

Temperature compensated, durable housing and minimal moving parts ensure reliability.

 

Highly Portable

All-in-one handheld instrument with long-life batteries that fully charge in less than 3 hours, an internal log-life pump, and one-hand operation.

 

Facility Monitoring

Easy to use software allows programming unique environmental profiles and recording sample data.

Common Applications

  • Industrial/Occupational Hygiene Area Surveys
  • Particulate Matter Studies
  • Immediate spot measurement
  • Aerosol Research Studies

 

Specification:

Measurement Method

Light scatter

Particle Size Screening

Standard Model: TSP

Deluxe Model: TSP, PM 2.5, or PM 10 Selective

Sample Method

Pump sample draw

Range

0.1 – 10,000 μg/m3 (0.0001 to 10 mg/ m3)

Sample Options

1 minute, Continuous, or 15 min STEL

Resolution

1 μg/m3 (0.001 mg/m3)

Sensitivity

1 μg/m3 (0.001 mg/m3)

Accuracy

± 5% traceable standard with 0.6 μm (0.0006 mg) PSL

Data logging

4000 records of STEL, Max, Min and average reading and k-factor

Power:

7.2V Lithium Ion self-contained battery pack providing 30 hours of typical intermittent operation and up to 12 hours continuous use

Charging Time

Less than 3 hours

Communications

USB Mini B Type

Dimension

9.5W x 17.2H x 5.1D cm

Weight

0.64 kg