Disable Preloader

Lorem ipsum is simply free text dolor sit am adipi we help you ensure everyone is in the right jobs sicing elit, sed do consulting firms Et leggings across the nation tempor.

Latest News

Contact Info
images

Ozone Switch [OS-4]

Acts like a thermostat to control ozone generators and alarms based on adjustable ozone concentration set-points. Ozone ppm readout via digital meter. Simple and effective. Ships with SM-7/SM-EC 0-20 ppm sensor module.

 

Features:

  • Easy to install. Minimal adjustments.
  • Sensor is in a replaceable precalibrated module. Separated by a cable, the instrument is not in the high ozone area.
  • Sensor reading is temperature and relative humidity compensated.
  • Replacement sensor is miniature, low cost, and pre-calibrated.
  • Operates from 12-24 VDC or from AC adapters available worldwide.
  • Alarms, data loggers, and other accessories available.
  • Programmable turn-on and turn-off set-points.
  • Indication when .1 ppm safety standard exceeded.

 

Specifications:

Range

03-20 ppm of ozone

Accuracy

10-15% (Depends on application)

Temperature and humidity

5-40°c 20-95% RH

Measurement principle

SPEC (Screen Printed Electrochemical) sensor

Size

127 × 64 × 42 mm (5 × 2.5 × 1.625 in)

Weight

 200 gr (7 oz)

Power requirements

12 VDC at 300 mA. AC adapters available worldwide

Output control

SPDT non-latching relay. Contacts rated at 5 amps, 250 VAC

.1 ppm danger alarm

Panel LED and SPDT relay 5 amp contacts

Connections

Terminal blocks for relay output and DC input. 2.5 mm DC power jack for AC adapter

Indicators

Green LED: instrument on. Yellow LED: >.1 ppm. Red LED: generator on

Outputs

0-2 V, 4-20mA, RS-232 digital

Case

Flame-resistant ABS with mounting ears