Ozone Calibration Source [306]

The 2B Technologies Model 306 Ozone Calibration Source™ is a portable source of ozone that allows you to calibrate any ozone monitor - not just those manufactured by 2B Tech.  The instrument scrubs ozone from ambient air and produces any mixing ratio of ozone in the range 30 to 1,000 ppbv.  The desired ozone concentration is chosen from the easy-to-use menu using a rotary select switch, or can be controlled remotely via the serial menu. The total output flow rate is 3.0 L/min volumetric, and the ozone mixing ratio is controlled so as to be independent of ambient temperature, pressure, and humidity.  You can attach the Ozone Calibration Source (OCS™) output directly to the inlet of any ozone monitor; the excess flow is vented through an ozone scrubber internal to the OCS. Although the instrument comes calibrated against a NIST-traceable ozone standard, the menu allows the user to change the calibration parameters of the ozone output if desired. For example, if you maintain an ozone reference instrument in the lab, you can use the OCS as a transfer standard for calibrating ozone instruments in the field relative to your lab standard.  The scheduler function allows you to schedule an autocalibration once daily that executes a series of up to 10 programmable output steps.

 

Specifications

Method of Ozone Production

UV Photolysis of O2 at 185 nm

Ozone Concentration Range

0 ppbv and 30 ppbv to 1,000 ppbv

Precision and Accuracy of  Output

Greater of 2 ppbv or 2% of ozone concentration

Response Time for Change in Ozone Output Concentration

30 s to reach 95% of concentration change

Flow Rate

3.0 Liter/min volumetric

Power Requirements

12 V dc or 120/240 V ac, 18 watt

Size

3.5 x 8.5 x 11 inches (9 x 21 x 29 cm)

Weight

5.6 lb (2.6 kg)

Analog Output

Yes; voltage output tracks ozone output and provides system error codes

Serial Connection

Yes; for remote operation and system diagnostics

Programmable Output

Yes; user-specified ozone output for up to 10 steps from 0 to 1,000 ppbv; stored in internal memory

Scheduler

Yes; once-daily execution of the step program at user-specified hour

Baud Rate

4800