Scrubberless Ozone Monitor [211]

The new Model 211 Scrubberless Ozone Monitor uses the proven ultraviolet (UV) absorption method in combination with our patented (US Patent No. 8,395,776 B2) gas-phase scrubber technology to measure ozone virtually free of interferences. The instrument is ideally suited for measurements of ozone in heavily polluted air where interference is likely from particulates, mercury, or VOCs. The Model 211 combines the stability of our proven dual beam ozone monitors with our selective scrubber technology to provide accurate ozone measurements in the most polluted air. The enhanced optical path length of 30 cm provides a precision of better than 0.5 ppb for 10-second measurements, and the instrument can output new ozone measurements as often as every 2 seconds. The Model 211 now comes with SD flash card memory as a standard feature.  The Model 211 Scrubberless Ozone Monitor has been designated by the EPA as a Federal Equivalent Method (FEM): EQOA–0514–215.

 

Features

  • Interference-free measurement of ambient ozone
  • DewLine™ for elimination of water vapor interference, a unique feature of 2B Tech instruments
  • Inexpensive, simple, and robust instrument design in a 19” rack mount enclosure
  • Internal data logger
  • SD flash memory for virtually unlimited portable data storage
  • Dual beam, long path (30 cm) optical bench with high precision (< 0.5 ppb)
  • NIST-traceable calibration (calibration suggested annually)
  • Internal long-life sample pump (20,000 hours)
  • EPA Federal Equivalent Method (FEM) for compliance monitoring (EQOA–0514–215)
  • Candidate to become a new EPA Federal Reference Method (FRM)
  • Convenient user interface, including remote operation via RS-232 communication
  • Internal NO generator from non-toxic, non-corrosive nitrous oxide (N2O)
  • Very low NO concentration used in comparison to ozone monitors based on NO chemiluminescence

 

Specifications

Measurement Principle

UV Absorbance at 254 nm; Dual Beam

Federal Equivalent Method (FEM)

Yes, EQOA–0514–215

Linear Dynamic Range

0-2,000 ppb (0-2 ppm)

Resolution

0.1 ppb

Precision (1σ rms noise)

Greater of 0.5 ppb or 1% of reading for 10-s average

Accuracy

Greater of 1.0 ppb or 2% of reading

Limit of Detection (2σ)

1.0 ppb

NIST-Traceable Calibration

Yes

Flow Rate (nominal)

~2 Liter/min

Flow Rate Requirement

>1.2 L/min

Baseline Drift

< 1 ppb/day
< 3 ppb/year

Sensitivity Drift

< 0.5%/day
< 3%/year

Response Time, 100% of Step Change

20 s for 10-s averaging; 4 s for 2-s measurement

Measurement and Averaging Times

2 s, 10 s, 1 min, 5 min, 1 hr

Internal Data Logger Capacity

16,383 lines (10-s avg. = 1.9 days; 1-min avg = 11 days; 5-min avg = 1.9 mo;
1-hr avg = 1.9 yr)

SD Card Logger Capacity

Minimum 2 GB (> 5-year capacity for 10-s measurement mode)

Ozone Units

ppb, ppm, pphm, µg m-3, mg m-3

Pressure Units

mbar, torr

Temperature Units

°C, K

T and P Corrected

Yes

Operating Temperature Range

10 to 50 °C

Operating Altitude Range

~0-13.5 km (150-1,013 mbar) standard; optional ~0-25 km (30-1,013 mbar) with upgraded pressure sensor

Power Requirement;
Supplied by Battery or 110/220 VAC Power Pack

11-14 V dc or 120/240 V ac, 1 A at 12 V, 12 watt

Size

Rackmount: 17" w x 14.5" d x 5.5" h (43 x 37 x 14 cm)

Weight

14.7 lb (6.7 kg)

Data Outputs

RS232, 0-2.5 V Analog, optional 4-20 mA Analog, optional USB adapter

Data Transfer Baud Rates

2400, 4800, 19200

Output Ranges

User-Defined Scaling Factor in Menu

DewLine™

Yes

Long-Life Pump

Yes, 20,000 hr

Flow Meter

Yes

Options

Battery; External Particle Filter; Bluetooth; External Serial-to-USB Adapter; 4-20 mA Current Output; Lamp Heater; High-Altitude Upgrade (pressure sensor)