The Model 205 Dual Beam Ozone Monitor makes use of two detection cells to improve precision, baseline stability, and response time. In the Dual Beam instrument, UV light intensity measurements Io (ozone-scrubbed air) and I (unscrubbed air) are made simultaneously. Combined with other improvements, this made it possible to reduce the time between ozone measurements to 2 seconds, making our instrument the fastest UV-based ozone monitor on the market, while still retaining the small size, weight, and power requirements of our popular Model 202 Ozone Monitor. Fast measurements are especially desirable for aircraft and balloon measurements where high spatial resolution is needed. For 10-s averaging, a precision of 1.0 ppb is achieved. Alternatively, one can choose to average more points internally to obtain even better precision. Flash card memory and a long-life, quiet internal air sampling pump are now standard on the instrument. The Model 205 Dual Beam Ozone Monitor has been approved by the EPA as a Federal Equivalent Method (FEM) based on a modification of the Model 202: EQOA-0410-190.
Features
Specifications
Measurement Principle | UV Absorption at 254 nm; Dual Beam |
Federal Equivalent Method (FEM) | Yes, EQOA-0410-190 |
Linear Dynamic Range | 0 to 250 ppm |
Resolution | 0.1 ppb |
Precision (1σ rms noise) | Greater of 1.0 ppb or 2% of reading for 10-s average |
Accuracy | Greater of 1.0 ppb or 2% of reading |
Limit of Detection (2σ) | 2.0 ppb for 10-s average |
NIST-Traceable Calibration | Yes |
Measurement Interval | 2 s (Data averaging options: 10 s, 1 min, 5 min, 1 hr) |
Flow Rate (nominal) | ~1.8 Liter/min |
Flow Rate Requirement | >1.2 L/min |
Baseline Drift | < 1 ppb/day |
Sensitivity Drift | < 1%/day |
Measurement Time, Frequency | 2 s, 0.5 Hz |
Response Time, 100% of Step Change | 4 s, 2 points |
Measurement and Averaging Times | 2 s, 10 s, 1 min, 5 min, 1 hr |
Ozone Units | ppb, pphm, ppm, µg m-3, mg m-3 |
Pressure Units | mbar, torr |
Temperature Units | °C, K |
T and P Corrected | Yes |
Operating Temperature Range | 0 to 50 °C standard; optional -20 to 50 °C with low temperature modifications (pump and lamp heater cold-weather package) |
Operating Altitude Range | ~0-13.5 km (150-1,013 mbar) standard; optional ~0-25 km (30-1,013 mbar) with upgraded pressure sensor |
Power Requirement; | 12 V dc or 120/240 V ac, nominally 650 mA at 12 V, ~8 watt (~7 watt with cell heater unplugged) |
Size | 3.7 x 8.5 x 11.5 inches (9.5 x 21.5 x 30 cm) |
Weight | 5.8 lb (2.6 kg); 3.1 lb (1.4 kg) without instrument case |
Internal Data Logger Capacity | 16,383 lines (2-s fast measurement mode = 9.1 hr; 10 s avg. = 1.9 days; |
SD Card Logger Capacity | > 2 GB (> 5-year capacity for 10-s averaging mode) |
Analog Inputs for Internal Logging of Other Instruments | 3 Analog Inputs, 0-2.5 V; For example, could log external T, P, and RH |
Data Outputs | RS232, 0-2.5 V Analog, LCD Display, External USB Converter (Optional 4-20 mA Current; request quote) |
Data Transfer Baud Rates | 2400, 4800, 19200 |
Output Ranges | User-Defined Scaling Factor in Menu |
DewLine™ | Yes |
Quiet, Long-Life Internal Air Pump | Yes |
Flow Meter | Yes |
Options | Battery; External Particle Filter; Serial-to-USB Adapter; GPS; Bluetooth; 4-20 mA Current Output; Cold-Weather Package (pump, lamp heater); High-Altitude Upgrade (pressure sensor); Extra-Long-Life, Low-Temperature Internal Air Pump; Low-Power Option; Rack-Mount Case Configuration |